Mouse MMP-3 (Matrix Metalloproteinase 3) ELISA Kit

Mouse MMP-3 (Matrix Metalloproteinase 3) ELISA Kit

Assay Range: 15.6-1000pg/mL
Sensitivity: 6.1pg/mL
Size: 96T Price: $420
Size: 48T Price: $336
SKU: EM20920 Category: ELISA Kits Tags: , , , , ,

Datasheet

KEY FEATURES

Full NameMatrix Metalloproteinase 3
SynonymMMP3, CHDS6, SL-1, STMY, STMY1, STR1, Progelatinase
Assay TypeSandwich
ReactivityMouse
Range15.6-1000pg/mL
Sensitivity6.1pg/mL
Sample TypeSerum, plasma and other biological fluids
Sample Volume100μL
Detection WavelengthOD450
Transportation Temperature2-8°C
SpecificityThe kit detected Mouse MMP-3 in the samples and no significant cross-species reactions were found
Microplate96-wells plate breakable into 12 x 8 wells strip

TEST PRINCIPLE

This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Mouse MMP-3. Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Mouse MMP-3 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Mouse MMP-3, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Mouse MMP-3. You can calculate the concentration of Mouse MMP-3 in the samples by comparing the OD of the samples to the standard curve.

ELISA KIT COMPONENTS

Upon receipt, unpack promptly and store as recommended in the instructions.

ComponentsSpecificationsStorage and Notes
Micro Plate
96T: 8 wells×12 strips
48T: 8 wells×6 strips
Unopened: -20°C, 12 months
Unused: Put it back in the aluminum foil bag and seal it, store it at -20°C.
Reference Standard
96T: 2 vials
48T: 1 vial
Unopened: -20°C, 12 months
Please use freshly dissolved standards for each experiment. Discard any unused standards after dissolution.
Biotinylated Detection Ab Concentrate (100×)
96T: 120μL×1 vial
48T: 60μL×1 vial
Unopened: -20°C, 12 months
Unused: Please seal the concentrate and store it at -20°C, and discard the working solution.
HRP Conjugate Concentrate (100×)
96T: 120μL×1 vial
48T: 60μL×1 vial
Unopened: -20°C(Protect from light), 12 months
Unused: Please seal the concentrate and store it at -20°C, and discard the working solution.
Biotinylated Detection Ab Diluent
14mL×1
2-8℃, 12 months
HRP Conjugate Diluent
14mL×1
2-8℃, 12 months
Reference Standard & Sample Diluent20mL×1
2-8℃, 12 months
Washing Buffer Concentrate (25×)30mL×1
2-8℃, 12 months
Substrate Reagent(TMB)10mL×12-8°C(Protect from light),12 months
Stop Solution7mL×12-8°C/Room temperature

ASSAY PROCEDURES

TYPICAL DATA

 Standard Curve Image

Mouse MMP-3 ELISA Standard Curve

Typical data are for reference only and curves should be replotted for each experiment. The Logistics function is recommended for fitting.

PRECISION

Intra-Assay Precision (Precision within an assay): Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-Assay Precision (Precision between assays): Three samples of known concentration were tested in twenty separate assays to assess inter-assay precision. Assays were performed by at least three technicians using two lots of components.

Intra-assay PrecisionInter-assay Precision
Sample123123
n202020202020
Mean (pg/mL)53.55122.33397.4148.82111.38368.98
Standard deviation35.5812.523.36.0814.21
CV(%)5.64.563.156.765.463.85

RECOVERY

The recovery of Mouse MMP-3 spiked to three different levels in samples throughout the range of the assay in various matrices was evaluated.

Sample TypeRange (%)Average Recovery (%)
Serum(n=8)94-107101
EDTA plasma (n=8)88-9994
Cell culture media (n=8)89-10194

LINEARITY

To assess the linearity of the assay, samples containing and/or spiked with high concentrations of Mouse MMP-3 in various matrices were diluted with the Reference Standard & Sample Diluent to produce samples with values within the dynamic range of the assay.