KEY FEATURES
Full Name | Proneurotensin/neuromedin N |
---|---|
Synonym | Proneurotensin 1-117 |
Assay Type | Competitive |
Reactivity | Human |
Range | 12.35-1000pg/mL |
Sensitivity | 4.84pg/mL |
Sample Type | Serum,plasma and other biological fluids |
Sample Volume | 50μL |
Detection Wavelength | OD450 |
Transportation Temperature | 2-8°C |
Specificity | The kit detected Human ProNT/NMN in the samples and no significant cross-species reactions were found |
Microplate | 96-wells plate breakable into 12 x 8 wells strip |
TEST PRINCIPLE
This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Human ProNT/NMN. During the reaction, Human ProNT/NMN in the sample or standard competes with a fixed amount of Human ProNT/NMN on the solid phase supporter for sites on the Biotinylated Detection Ab specific to Human ProNT/NMN. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of Human ProNT/NMN in tested samples can be calculated by comparing the OD of the samples to the standard curve. |
ELISA KIT COMPONENTS
Upon receipt, unpack promptly and store as recommended in the instructions.
Components | Specifications | Storage and Notes |
---|---|---|
Micro Plate | 96T: 8 wells×12 strips 48T: 8 wells×6 strips | Unopened: -20°C, 12 months Unused: Put it back in the aluminum foil bag and seal it, store it at -20°C. |
Reference Standard | 96T: 2 vials 48T: 1 vial | Unopened: -20°C, 12 months Please use freshly dissolved standards for each experiment. Discard any unused standards after dissolution. |
Biotinylated Detection Ab Concentrate (100×) | 96T: 120μL×1 vial 48T: 60μL×1 vial | Unopened: -20°C, 12 months Unused: Please seal the concentrate and store it at -20°C, and discard the working solution. |
HRP Conjugate Concentrate (100×) | 96T: 120μL×1 vial 48T: 60μL×1 vial | Unopened: -20°C(Protect from light), 12 months Unused: Please seal the concentrate and store it at -20°C, and discard the working solution. |
Biotinylated Detection Ab Diluent | 14mL×1 | 2-8℃, 12 months |
HRP Conjugate Diluent | 14mL×1 | 2-8℃, 12 months |
Reference Standard & Sample Diluent | 20mL×1 | 2-8℃, 12 months |
Washing Buffer Concentrate (25×) | 30mL×1 | 2-8℃, 12 months |
Substrate Reagent(TMB) | 10mL×1 | 2-8°C(Protect from light),12 months |
Stop Solution | 7mL×1 | 2-8°C/Room temperature |
ASSAY PROCEDURES

TYPICAL DATA

Human ProNT/NMN ELISA Standard Curve
Typical data are for reference only and curves should be replotted for each experiment. The Logistics function is recommended for fitting.
PRECISION
Intra-Assay Precision (Precision within an assay): Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays): Three samples of known concentration were tested in twenty separate assays to assess inter-assay precision. Assays were performed by at least three technicians using two lots of components.
Intra-assay Precision | Inter-assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 52.58 | 104.08 | 410.45 | 49.02 | 100.54 | 402.37 |
Standard deviation | 2.83 | 4.49 | 14.98 | 3.25 | 5.31 | 16.05 |
CV(%) | 5.39 | 4.31 | 3.65 | 6.62 | 5.28 | 3.99 |
RECOVERY
The recovery of Human ProNT/NMN spiked to three different levels in samples throughout the range of the assay in various matrices was evaluated.
Sample Type | Range (%) | Average Recovery (%) |
---|---|---|
Serum(n=8) | 88-104 | 95 |
EDTA plasma (n=8) | 90-105 | 97 |
Cell culture media (n=8) | 87-99 | 93 |
LINEARITY
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of Human ProNT/NMN in various matrices were diluted with the Reference Standard & Sample Diluent to produce samples with values within the dynamic range of the assay.
