Human TAP (Trypsinogen Activation Peptide) ELISA Kit

Human TAP (Trypsinogen Activation Peptide) ELISA Kit

Assay Range: 0.124-10ng/mL
Sensitivity: 0.054ng/mL
Size: 96T Price: $520
Size: 48T Price: $416
SKU: EH24409 Category: ELISA Kits Tags: ,

KEY FEATURES

Full NameTrypsinogen Activation Peptide
SynonymTAP
Assay TypeCompetitive
ReactivityHuman
Range0.124-10ng/mL
Sensitivity0.054ng/mL
Sample TypeSerum,plasma and other biological fluids
Sample Volume50μL
Detection WavelengthOD450
Transportation Temperature2-8°C
SpecificityThe kit detected Human TAP in the samples and no significant cross-species reactions were found
Microplate96-wells plate breakable into 12 x 8 wells strip

TEST PRINCIPLE

This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Human TAP. During the reaction, Human TAP in the sample or standard competes with a fixed amount of Human TAP on the solid phase supporter for sites on the Biotinylated Detection Ab specific to Human TAP. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of Human TAP in tested samples can be calculated by comparing the OD of the samples to the standard curve.

ELISA KIT COMPONENTS

Upon receipt, unpack promptly and store as recommended in the instructions.

ComponentsSpecificationsStorage and Notes
Micro Plate
96T: 8 wells×12 strips
48T: 8 wells×6 strips
Unopened: -20°C, 12 months
Unused: Put it back in the aluminum foil bag and seal it, store it at -20°C.
Reference Standard
96T: 2 vials
48T: 1 vial
Unopened: -20°C, 12 months
Please use freshly dissolved standards for each experiment. Discard any unused standards after dissolution.
Biotinylated Detection Ab Concentrate (100×)
96T: 120μL×1 vial
48T: 60μL×1 vial
Unopened: -20°C, 12 months
Unused: Please seal the concentrate and store it at -20°C, and discard the working solution.
HRP Conjugate Concentrate (100×)
96T: 120μL×1 vial
48T: 60μL×1 vial
Unopened: -20°C(Protect from light), 12 months
Unused: Please seal the concentrate and store it at -20°C, and discard the working solution.
Biotinylated Detection Ab Diluent
14mL×1
2-8℃, 12 months
HRP Conjugate Diluent
14mL×1
2-8℃, 12 months
Reference Standard & Sample Diluent20mL×1
2-8℃, 12 months
Washing Buffer Concentrate (25×)30mL×1
2-8℃, 12 months
Substrate Reagent(TMB)10mL×12-8°C(Protect from light),12 months
Stop Solution7mL×12-8°C/Room temperature

ASSAY PROCEDURES

TYPICAL DATA

 Standard Curve Image

Human TAP ELISA Standard Curve

Typical data are for reference only and curves should be replotted for each experiment. The Logistics function is recommended for fitting.

PRECISION

Intra-Assay Precision (Precision within an assay): Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-Assay Precision (Precision between assays): Three samples of known concentration were tested in twenty separate assays to assess inter-assay precision. Assays were performed by at least three technicians using two lots of components.

Intra-assay PrecisionInter-assay Precision
Sample123123
n202020202020
Mean (ng/mL)0.471.494.40.481.44.64
Standard deviation0.030.090.150.030.080.17
CV(%)5.875.743.316.715.593.68

RECOVERY

The recovery of Human TAP spiked to three different levels in samples throughout the range of the assay in various matrices was evaluated.

Sample TypeRange (%)Average Recovery (%)
Serum(n=8)89-10495
EDTA plasma (n=8)87-9992
Cell culture media (n=8)89-10194

LINEARITY

To assess the linearity of the assay, samples containing and/or spiked with high concentrations of Human TAP in various matrices were diluted with the Reference Standard & Sample Diluent to produce samples with values within the dynamic range of the assay.

Pathways associated with ELISA Kit

Activation Of Matrix Metalloproteinases Pathway Activation Of Matrix Metalloproteinases Pathway
Cobalamin (Cbl, Vitamin B12) Transport And Metabolism Pathway Cobalamin (Cbl, Vitamin B12) Transport And Metabolism Pathway
Defective AMN Causes Hereditary Megaloblastic Anemia 1 Pathway Defective AMN Causes Hereditary Megaloblastic Anemia 1 Pathway
Defective BTD Causes Biotidinase Deficiency Pathway Defective BTD Causes Biotidinase Deficiency Pathway
Defective CD320 Causes Methylmalonic Aciduria Pathway Defective CD320 Causes Methylmalonic Aciduria Pathway
Defective CUBN Causes Hereditary Megaloblastic Anemia 1 Pathway Defective CUBN Causes Hereditary Megaloblastic Anemia 1 Pathway
Defective GIF Causes Intrinsic Factor Deficiency Pathway Defective GIF Causes Intrinsic Factor Deficiency Pathway
Defective HLCS Causes Multiple Carboxylase Deficiency Pathway Defective HLCS Causes Multiple Carboxylase Deficiency Pathway
Defective LMBRD1 Causes Methylmalonic Aciduria And Homocystinuria Type CblF Pathway Defective LMBRD1 Causes Methylmalonic Aciduria And Homocystinuria Type CblF Pathway
Defective MMAA Causes Methylmalonic Aciduria Type CblA Pathway Defective MMAA Causes Methylmalonic Aciduria Type CblA Pathway