KEY FEATURES
Full Name | Amyloid Beta Peptide 1-40 |
---|---|
Synonym | Ab1-40 |
Assay Type | Competitive |
Reactivity | Mouse |
Range | 12.35-1000pg/mL |
Sensitivity | 5.38pg/mL |
Sample Type | Serum,plasma and other biological fluids |
Sample Volume | 50μL |
Detection Wavelength | OD450 |
Transportation Temperature | 2-8°C |
Specificity | The kit detected Mouse Ab1-40 in the samples and no significant cross-species reactions were found |
Microplate | 96-wells plate breakable into 12 x 8 wells strip |
TEST PRINCIPLE
This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Mouse Ab1-40. During the reaction, Mouse Ab1-40 in the sample or standard competes with a fixed amount of Mouse Ab1-40 on the solid phase supporter for sites on the Biotinylated Detection Ab specific to Mouse Ab1-40. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of Mouse Ab1-40 in tested samples can be calculated by comparing the OD of the samples to the standard curve. |
ELISA KIT COMPONENTS
Upon receipt, unpack promptly and store as recommended in the instructions.
Components | Specifications | Storage and Notes |
---|---|---|
Micro Plate | 96T: 8 wells×12 strips 48T: 8 wells×6 strips | Unopened: -20°C, 12 months Unused: Put it back in the aluminum foil bag and seal it, store it at -20°C. |
Reference Standard | 96T: 2 vials 48T: 1 vial | Unopened: -20°C, 12 months Please use freshly dissolved standards for each experiment. Discard any unused standards after dissolution. |
Biotinylated Detection Ab Concentrate (100×) | 96T: 120μL×1 vial 48T: 60μL×1 vial | Unopened: -20°C, 12 months Unused: Please seal the concentrate and store it at -20°C, and discard the working solution. |
HRP Conjugate Concentrate (100×) | 96T: 120μL×1 vial 48T: 60μL×1 vial | Unopened: -20°C(Protect from light), 12 months Unused: Please seal the concentrate and store it at -20°C, and discard the working solution. |
Biotinylated Detection Ab Diluent | 14mL×1 | 2-8℃, 12 months |
HRP Conjugate Diluent | 14mL×1 | 2-8℃, 12 months |
Reference Standard & Sample Diluent | 20mL×1 | 2-8℃, 12 months |
Washing Buffer Concentrate (25×) | 30mL×1 | 2-8℃, 12 months |
Substrate Reagent(TMB) | 10mL×1 | 2-8°C(Protect from light),12 months |
Stop Solution | 7mL×1 | 2-8°C/Room temperature |
ASSAY PROCEDURES

TYPICAL DATA

Mouse Ab1-40 ELISA Standard Curve
Typical data are for reference only and curves should be replotted for each experiment. The Logistics function is recommended for fitting.
PRECISION
Intra-Assay Precision (Precision within an assay): Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays): Three samples of known concentration were tested in twenty separate assays to assess inter-assay precision. Assays were performed by at least three technicians using two lots of components.
Intra-assay Precision | Inter-assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 51.84 | 94.28 | 468.5 | 55.33 | 100.89 | 428.44 |
Standard deviation | 3.44 | 5.53 | 17.8 | 2.96 | 5.51 | 14.01 |
CV(%) | 6.64 | 5.87 | 3.8 | 5.35 | 5.46 | 3.27 |
RECOVERY
The recovery of Mouse Ab1-40 spiked to three different levels in samples throughout the range of the assay in various matrices was evaluated.
Sample Type | Range (%) | Average Recovery (%) |
---|---|---|
Serum(n=8) | 88-100 | 94 |
EDTA plasma (n=8) | 93-110 | 101 |
Cell culture media (n=8) | 94-106 | 99 |
LINEARITY
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of Mouse Ab1-40 in various matrices were diluted with the Reference Standard & Sample Diluent to produce samples with values within the dynamic range of the assay.