Nitrofurantoin (AHD) ELISA Kit

Nitrofurantoin (AHD) ELISA Kit

Cat: EKF1032

Assay Range: 0.02 ppb~1.62 ppb (ng/mL)

Sensitivity: 0.02 ppb (ng/mL)

Size: 96T    Price: $320

Sample type: Muscle, Liver, Honey, Milk, Milk powder, Egg powder ,Feed, Egg, Cooked food

Synonym: 1-Aminohydantoin

Category: Food Safety Test Kits Tags: , , , , ,

Product description

This ELISA kit is rapid, quantitative enzyme-linked immunosorbent assays (ELISA) for the determination of Nitrofurantoin content in Food samples.


Key features

Assay Range:   0.02 ppb~1.62 ppb

Sensitivity:   0.02 ppb (ng/mL)

Species reactivity:   Universal

Assay time:   60min

Result type:   Quantitative

Category:   Residues&contaminants

Kit type:   ELISA

Detection type:   Competitive

Precision:   CV<10%

Recovery rate: 

Muscle, liver, egg

85±25%





Honey, milk

80±20%





Milk powder, egg powder, feed

85±25%





Cooked food

80±20%





Specificity:

AHD

100%





AMOZ, AOZ, SEM

<0.1%





Detection limits:

Muscle, liver, egg

0.04 ppb





Fish, shrimp

0.06 ppb





Honey, milk

0.04 ppb





Milk powder, egg powder, feed

0.04 ppb





Cooked food

0.04 ppb






Storage and shelf life

The kit should be stored at 2~8°C and must not be frozen. The shelf life of this product is 1 year. Do not use the kit beyond the expiration date.


Test principle

This assay employs the competitive inhibition enzyme immunoassay technique. The kit consists of an enzyme-labeled plate pre-coated with coupled antigens, horseradish enzyme markers, antibodies, standard substances, and other supporting reagents. During detection, standard substances or sample solutions are added. Nitrofurantoin in the samples competes with the pre-coated coupled antigens on the enzyme-labeled plate for Nitrofurantoin specific antibody. After adding the enzyme markers, TMB substrate is used for color development. The absorbance value of the sample is negatively correlated with the content of Nitrofurantoin it contains, and the residual amount of Nitrofurantoin in the sample can be obtained by comparing with the standard curve. 


Materials supplied

Item

Quantity





Assay Plate

96T





Standard

7 x 1 mL





Derivatization reagent

1 x 10 mL





HRP Conjugate

1 x 5.5 mL





Antibody

1 x 5.5 mL





TMB Substrate Reagent (A\B)

2 x 6 mL





Stop Solution

1 x 6 mL





Wash Buffer (20×)

1 x 40 mL





Reconstitution Buffer (2×)

1 x 50 mL





Plate Sealer

1 piece





Instruction Manual

1 piece





Sealed Bag

1 copy






Standard curve reference

Standard

1

2

3

4

5

6




Concentration

(ppb=ng/mL=ng/g)

0

0.02

0.06

0.18

0.54

1.62




OD Standard

2.305

1.836

1.408

0.886

0.495

0.246