EnkiLife · Cat. No. EH27095T · 96 tests
The EnkiLife Human IL‑12 (Interleukin 12) ELISA Kit measures human interleukin‑12 by colorimetric sandwich ELISA. EnkiLife evaluated spike recovery in serum, EDTA plasma, and cell culture media.
For research use only.
Plate layout shown for 12 removable strips with 8 wells each. Reserve wells for calibration and controls when planning sample capacity.
- Assay range
- 15.6–1,000 pg/mL
- Reported sensitivity
- 5.7 pg/mL
- Sample per well
- 100 µL
- Absorbance readout
- 450 nm
Assay specifications
| Attribute | Specification |
|---|---|
| Target and species | Human interleukin‑12 (IL‑12) |
| Assay format | Quantitative, colorimetric sandwich enzyme-linked immunosorbent assay (ELISA) |
| Sample matrices evaluated | Serum, EDTA plasma, and cell culture media in manufacturer spike-recovery studies |
| Reported assay range | 15.6–1,000 pg/mL |
| Reported sensitivity | 5.7 pg/mL |
| Sample volume | 100 µL per sample well |
| Detection instrument | Microplate reader configured for absorbance at 450 nm |
| Pack format | 96 tests in a pre-coated strip plate. 12 strips × 8 wells. |
Use the reported assay range when planning quantitation. Ask technical support for the sensitivity method or a documented lower limit of quantitation if your study requires one.
Sample compatibility and preparation
Choose handling and dilution conditions for your sample matrix. Apply the preparation instructions in the insert supplied with your kit lot.
| Matrix | Published evaluation | Before loading the plate |
|---|---|---|
| Serum | Spike recovery evaluated. | Separate serum from cellular material and log sample handling and any dilution. |
| EDTA plasma | Spike recovery evaluated using EDTA plasma. | Record the anticoagulant. Ask technical support about citrate or heparin plasma before relying on either matrix. |
| Cell culture media | Spike recovery evaluated. | Follow the instructions for clarified culture supernatant. Record the medium and any additives that could affect recovery. |
Plan a matrix-specific check for other fluids or extracts. Follow the lot insert for sample storage, aliquoting, and handling of hemolyzed or lipemic samples. Read the general ELISA sample-preparation guide.
Typical data and validation
What EnkiLife evaluated
- Within-run precision
- Three concentration levels, each measured 20 times on one plate.
- Between-run precision
- Three concentration levels across 20 assays, with at least three operators and two component lots.
- Spike recovery
- Three spike levels assessed in serum, EDTA plasma, and cell culture media.
- Dilution linearity
- High-concentration samples were diluted with the kit’s standard and sample diluent to assess performance within the assay range.
For validation values associated with your lot, contact technical support with the catalog number and lot number.
Sandwich ELISA workflow
The pre-coated antibody captures human IL‑12. A biotinylated detection antibody and avidin–horseradish peroxidase (HRP) conjugate generate a color signal with tetramethylbenzidine (TMB). Absorbance rises with IL‑12 concentration across the working range.
- Capture
Load the dilution series, diluent blank, and samples. Reserve replicate wells for the calibration standards.
- Detect
Add the detection antibody and then the HRP conjugate. Apply the wash steps specified for your lot.
- Develop and read
Develop with TMB, stop the reaction and measure absorbance at 450 nm.
- Quantify
Fit your calibration data and interpolate the sample concentrations.
Lab equipment and consumables
Have a 450 nm microplate reader, a temperature-controlled incubator, calibrated pipettes and disposable tips, sample tubes, absorbent paper, and distilled or deionized water available.
Use the kit insert for exact instrument setup, reagent preparation, wash settings, incubation time, and temperature.
Calculate sample concentration
Plot mean standard absorbance against concentration and fit a four-parameter logistic (4PL) curve as described in the manual. Apply the blank handling specified in your lot insert and check replicate agreement before interpolation.
Original sample concentration = interpolated concentration × pre-assay dilution factor.
If a result is above the upper standard, remeasure it with an appropriate dilution under the lot instructions. Apply your lab’s reporting policy to values below the working range.
96-test kit components and handling
| Component | Role or confirmed format |
|---|---|
| Pre-coated microplate | 12 × 8-well strips for antigen capture. |
| Human IL‑12 reference standard | Two vials of standard powder for preparing a calibration series. Reconstitute the standard according to the lot insert. |
| Biotinylated detection antibody concentrate | 100× concentrate with its own working-solution diluent. |
| HRP conjugate concentrate | 100× concentrate with its own working-solution diluent. |
| Standard and sample diluent | One 20 mL container used for calibration and sample dilution. |
| Wash buffer concentrate | Prepare the working wash buffer according to the bottle label and lot insert. |
| TMB substrate and stop solution | Reagents for color development and reaction termination. |
See equipment and consumables to prepare separately.
Shipping and storage
EnkiLife lists a shipping temperature of 2–8 °C. Unpack on receipt and check each component’s storage instructions. The manufacturer specifies −20 °C for unopened plate strips, reference standard, and the concentrated detection reagents, with light protection for the HRP conjugate and TMB as indicated on the insert. Reseal unused strips and prepare fresh standard working solutions for each run.
Product documents and technical support
EH27095 instructions
Download English datasheet (PDF, 13 pages)
About 380 KiB. The public guide covers assay principle, sample preparation, procedure and data analysis. Use the insert provided with your kit for lot-specific instructions.
EH27095 safety data sheet
Download English SDS (PDF, 16 pages)
Revision 3, September 2025. About 856 KiB. Review its handling and disposal information before working with the kit.
For a certificate of analysis or a replacement lot insert, request documents for EH27095 and include the lot number shown on the box.
Troubleshooting
| Observation | First checks |
|---|---|
| Poor standard curve | Check standard reconstitution and dilution order, pipette calibration, and the correct plate map. |
| Weak color or low signal | Verify reagent preparation, incubation settings from the insert, and the 450 nm reader setup. |
| High background | Check the blank wells and the working wash buffer. Verify complete aspiration and open washer ports if using an automated washer. |
| High variation between replicates | Check delivered volumes, consistency of timing, and whether any wells were left partially aspirated. |
| Sample above the curve | Select a suitable additional dilution using your lot instructions, repeat the measurement, and record the total dilution factor. |
For further help, send the kit lot, sample matrix, dilution, plate map, and raw reader output to EnkiLife technical support. The general ELISA troubleshooting guide covers additional symptoms.
IL‑12 target background
Interleukin‑12 is a cytokine studied in innate and adaptive immunity. The intact IL‑12 cytokine comprises the p35 subunit encoded by IL12A (UniProt P29459) and the p40 subunit encoded by IL12B (UniProt P29460). IL‑23 also contains p40. If your experiment requires p70-specific measurement or separation from free p40 or IL‑23, request EH27095 specificity documentation before selecting the assay.
- Trinchieri G. Interleukin-12 and the regulation of innate resistance and adaptive immunity. Nature Reviews Immunology, 2003. PMID 12563297.
- Oppmann B et al. Novel p19 protein engages IL-12p40 to form a cytokine, IL-23, with biological activities similar as well as distinct from IL-12. Immunity, 2000. PMID 11114383.
Human protein identifiers. IL12A, Gene ID 3592 and RefSeq NP_000873.2. IL12B, Gene ID 3593 and RefSeq NP_002178.2.
Cite this reagent
Human IL-12 (Interleukin 12) ELISA Kit, EnkiLife, catalog no. EH27095.Record your kit lot in the methods section for traceability. Permanent product link.
Frequently asked questions
- Can I use EDTA plasma or cell culture media with EH27095?
- Yes. EnkiLife reports spike-recovery evaluations for EDTA plasma, cell culture media, and serum. Use the lot-specific preparation and dilution instructions for your sample type.
- Can I use heparin plasma, citrate plasma, or a different biological fluid?
- Ask technical support for matrix-specific guidance. Plan a recovery and dilution check before interpreting quantitative results for a matrix outside the documented recovery panel.
- How much sample is needed?
- Load 100 µL per sample well. Reserve additional volume for replicates and planned dilutions when calculating the total sample required.
- How should I use the 5.7 pg/mL sensitivity value?
- Treat 5.7 pg/mL as the manufacturer’s reported sensitivity. Plan quantitative analysis around the stated 15.6–1,000 pg/mL assay range and the standards measured in your own run.
- What if a sample gives a signal above the highest standard?
- Use the insert’s dilution guidance and rerun the sample within the working range. Apply the total pre-assay dilution factor to the concentration obtained from the new curve.
- Where can I confirm p70 specificity or cross-reactivity with p40?
- Ask EnkiLife technical support for assay-specific documentation and tell them which molecular forms and cross-reactants your experiment needs to distinguish.
- Which wash buffer instructions should I follow?
- Use the concentration on the bottle and the preparation directions in the insert supplied with your kit. Request lot-specific instructions if the labels or documents differ.
- How do I obtain a quote or the lot’s certificate?
- For a quote, email the order team with EH27095, the 96T pack, quantity, and destination. For a certificate of analysis, contact technical support with the kit lot number.
For questions that apply across kits, see EnkiLife’s general sandwich ELISA FAQ.
Order EH27095 or discuss your application
Other EnkiLife contact options · Browse human ELISA kits · Custom ELISA services